Immunofluorescence Analyzer

The RaFIA Immunofluorescence  Analyzer is a fluorescence immunoassay analyzing instrument intended for use by healthcare professionals to aid in the diagnosis of conditions such as cardiovascular disease, pregnancy, infection, diabetes, renal injury and cancer.

This analyzer uses an LED as the excitation light source. The emitted light from the fluorescence dye is collected and converted into an electrical signal. The signal is closely related to the amount of fluorescence dye molecules presented on the spot under examination.
After a buffer-mixed sample is applied to the test device, the test device is inserted into the analyzer and the concentration of the analyte is calculated by a pre-programmed calibration process. The RaFIA Immunofluorescence Analyzer can only accept test devices that are designed especially for this equipment.

This instrument provides reliable and quantitative results for a variety of analytes in human blood and urine within 20 minutes.

This instrument is for in vitro diagnostic use only. Any use or interpretation of preliminary test results must also rely on other clinical findings and the professional judgment of health care providers. Alternative test method(s) should be considered to confirm the test results obtained by this device.

ACCURACY:

  • Less than 3% intra-run CV
  • Eliminates subjective interpretation by operator
  • Results comparable to lab-based CLIA analyzer

EASE-OF-USE:

  • System uses an SD card to ensure the proper test is being run
  • Able to run whole blood, serum, plasma and urine specimens
  • Touch screen interface for easy operation
  • Lightweight (2.9 kg) and portable

DATA MANAGEMENT:

  • Results presented on screen or via the built-in printer
  • Storage capacity of 10,000 test results
SPECIFICATIONS
Human Machine Interface:4.3-inch LCD touch screen
Dimensions:280(L)×256(W)×102(H)mm
Weight:2.9 kg
Power Adapter:100-240 VAC, 50-60Hz
Data output:On screen/Printer
OPERATION ENVIRONMENT
Temperature:10~30℃
Humidity:≤70%
Atmosphere pressure:85.0~106.0 Kpa
Operation Altitude:<2000m
Location:Dry, clean, flat, horizontal surface, away from direct sunlight, mechanical vibration, and strong electromagnetic interference.
STORAGE ENVIRONMENTAL
Temperature:0~50℃
Humidity:≤93%
Storage Elevation:0~2000m
OPTICAL DESCRIPTION
Excitation Light Source:LED 365nm
Detector:Silicon photodiode

Catalog #

Product

FIA-100

Immunofluorescence Analyzer

Other products you might be interested in:

Catalog #

Product

FIA-101

Immunofluorescence Incubator

FOR EXPORT ONLY