Working range: 0.1–10.0 mg/L

Compatible with human plasma and whole blood specimens

Test results in 3 minutes!

RaFIA
®
D-Dimer FIA Test
Orange CE Mark
Cat. # F3010
The D-Dimer Fluorescence ImmunoAssay (FIA) Test is a lateral flow chromatographic fluorescence immunoassay for the quantitative detection of D-Dimer in human plasma or whole blood using the RaFIA system. This test is intended for in vitro diagnostic use only. D-Dimer is the fibrinolysis-specific degradation product found in human blood. D-Dimer is not normally present in human blood, except when the coagulation system has been activated, for instance because of the presence of thrombosis or disseminated intravascular coagulation. Therefore, D-Dimer is commonly used to exclude thromboembolic disease where the probability is low. While a negative result may rule out thrombosis, a positive result can indicate thrombosis but does not rule out other potential causes. In addition, D-Dimer can be used in the diagnosis of blood disorders disseminated by intravascular coagulation. Myocardial infarction, cerebral infarction, pulmonary embolism, venous thrombosis, surgery, cancer, disseminated intravascular coagulation, infection and tissue necrosis may lead to elevated concentration of D-Dimer. Increased concentration of D-Dimer may also be observed in the elderly or hospitalized patients with abnormal coagulation caused by bacteremia. The D-Dimer FIA Test was developed to detect D-Dimer in human plasma or whole blood within 3 minutes, without the use of cumbersome laboratory equipment, and requiring minimally trained laboratory personnel.
The D-Dimer Fluorescence ImmunoAssay (FIA) Test is a lateral flow chromatographic fluorescence immunoassay for the quantitative detection of D-Dimer in human plasma or whole blood using the RaFIA system. This test is intended for in vitro diagnostic use only. D-Dimer is the fibrinolysis-specific degradation product found in human blood. D-Dimer is not normally present in human blood, except when the coagulation system has been activated, for instance because of the presence of thrombosis or disseminated intravascular coagulation. Therefore, D-Dimer is commonly used to exclude thromboembolic disease where the probability is low. While a negative result may rule out thrombosis, a positive result can indicate thrombosis but does not rule out other potential causes. In addition, D-Dimer can be used in the diagnosis of blood disorders disseminated by intravascular coagulation. Myocardial infarction, cerebral infarction, pulmonary embolism, venous thrombosis, surgery, cancer, disseminated intravascular coagulation, infection and tissue necrosis may lead to elevated concentration of D-Dimer. Increased concentration of D-Dimer may also be observed in the elderly or hospitalized patients with abnormal coagulation caused by bacteremia. The D-Dimer FIA Test was developed to detect D-Dimer in human plasma or whole blood within 3 minutes, without the use of cumbersome laboratory equipment, and requiring minimally trained laboratory personnel.
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests

Working Range: 0.1–10.0 mg/L

 Intra-lot Precision: Determined by testing of D-Dimer reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision: Determined by testing of D-Dimer reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.

Accuracy: D-Dimer control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.

Linearity: A serial concentration of D-Dimer reference materials at 0.1–10.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

Helpful Documents

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests

Working Range: 0.1–10.0 mg/L

 Intra-lot Precision: Determined by testing of D-Dimer reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision: Determined by testing of D-Dimer reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.

Accuracy: D-Dimer control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.

Linearity: A serial concentration of D-Dimer reference materials at 0.1–10.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

Helpful Documents

YOU MAY ALSO BE INTERESTED IN
YOU MAY ALSO BE INTERESTED IN
RaFIA® cTnI FIA Test CE
Cat. # F3002
LEARN MORE
RaFIA® CK-MB FIA Test CE
Cat. # F3020
LEARN MORE
RaFIA® MYO FIA Test CE
Cat. # F3030
LEARN MORE
RaFIA® NT-proBNP FIA Test CE
Cat. # F3050
LEARN MORE
RaFIA® Immunofluorescence Analyzer
Cat. # FIA-100
LEARN MORE
RaFIA® Immunofluorescence Incubator
Cat. # FIA-101
LEARN MORE
RaFIA® cTnI FIA Test CE
RaFIA® CK-MB FIA Test CE
RaFIA® MYO FIA Test CE
RaFIA® NT-proBNP FIA Test CE